How To Troubleshoot Issues With Yingtai Laboratory Centrifuges During Use

Views: 156     Author: 英泰仪器有限公司     Publish Time: 2024-07-16      Origin: 长沙英泰仪器有限公司公众号

How to Troubleshoot Issues with Yingtai Laboratory Centrifuges During Use


Yingtai laboratory centrifuges primarily utilize the powerful centrifugal force generated by the high-speed rotation of the rotor to rapidly sediment particles in liquids, thereby separating substances with different sedimentation coefficients and buoyant densities within the sample. They play a crucial role in fields such as pharmaceuticals and biomedicine.


If issues arise with the laboratory centrifuge, how can we troubleshoot them ourselves? Firstly, remind users that when repairing the centrifuge, they must ensure that the power switch is turned off and the centrifuge is disconnected from the main power supply. Otherwise, opening the device casing might expose components that are live or pose a risk of personal injury. When it is necessary to open the casing for internal maintenance, the operator must have professional training to avoid damaging the equipment or causing personal safety incidents.


The following are simple troubleshooting steps for safety protection faults in laboratory centrifuges:

1. Imbalance: Check if the weight of the samples in the centrifuge is within the error range, and check if the imbalance socket on the control board is loose, etc.

2. Overspeed: Check if the speed sensor socket on the control board is loose or if the speed sensor is malfunctioning.

3. Overheating: Check if the compressor is working properly, if the temperature sensor socket on the control board is loose or if the temperature sensor is damaged, and if the temperature control solenoid valve socket is loose or if the corresponding solenoid valve is faulty.

4. Door Open: Confirm that the door of the equipment is closed, check if the socket on the control board is loose or if the door detection switch is damaged.

These checks can better help users accurately identify faults and conduct daily maintenance on the laboratory centrifuge, thereby extending its service life. If users are unable to identify the abnormality


Your cookie settings.

Personalised experiences at full control.

This website uses cookies and similar technologies (“cookies”). Subject to your consent, will use analytical cookies to track which content interests you, and marketing cookies to display interest-based advertising. We use third-party providers for these measures, who may also use the data for their own purposes. 

You give your consent by clicking "Accept all" or by applying your individual settings. Your data may then also be processed in third countries outside the EU, such as the US, which do not have a corresponding level of data protection and where, in particular, access by local authorities may not be effectively prevented. You can revoke your consent with immediate effect at any time. If you click on "Reject all", only strictly necessary cookies will be used.