Views: 237 Author: Site Editor Publish Time: 2024-07-22 Origin: Site
Yingtai: Low-Speed Centrifuge Routine Maintenance Guide
I. Common Issues and Troubleshooting for Low-Speed Centrifuges
1. No Power Indicator after Startup:
a) Check if the indoor power supply is normal.
b) Ensure the plug and socket are making good contact.
c) Inspect whether the fuse is blown.
2. Excessive Vibration after Startup:
a) Verify if there is significant weight deviation of the separation liquid when symmetrically loaded into the centrifuge tube.
b) Check if the rotor groove aligns with the motor shaft.
II. Installation of Yingtai Low-Speed Centrifuge Main Unit
1. Site Requirements:
a) The installation tabletop should be sturdy, indoors should be dry, clean, and avoid direct sunlight.
b) There should be an independent ground wire indoors to ensure electrical safety.
2. Installation of Tabletop Low-Speed Centrifuge Main Unit:
a) Securely mount the main unit onto a solid tabletop, ensuring all four feet are in contact with the surface.
b) Maintain a gap of more than 0.1m between the centrifuge and surrounding walls to ensure adequate ventilation.
c) AC power supply of 220V 50Hz, with wire cross-sectional area not less than 2.5 square millimeters, and the protective grounding line must be securely connected.