Yingtai: The Influence of Steam Dryness on Moist Heat Sterilization

Views: 412     Author: Site Editor     Publish Time: 2025-08-11      Origin: Site

Yingtai: The Influence of Steam Dryness on Moist Heat Sterilization  

 

Steam dryness (the degree to which steam is free of liquid water) is one of the core parameters affecting the efficacy of moist heat sterilization, directly influencing sterilization efficiency, penetration capability, and equipment stability.  

 

1. Heat Transfer Efficiency and Sterilization Speed  

Latent Heat Advantage of Dry Steam  

Saturated steam (dryness 100%) releases latent heat (approximately 529 cal/g of water) upon condensation, rapidly raising the temperature of sterilized items and accelerating microbial protein denaturation. Wet saturated steam (dryness <95%) contains liquid water, reducing latent heat release and decreasing heat transfer efficiency by 30%-50%, necessitating extended sterilization times or higher temperatures.  

 

Penetration Capability Comparison  

Dry steam penetrates porous materials (e.g., fabrics, instrument crevices) 2-3 times more efficiently than wet steam, ensuring uniform heat distribution and avoiding localized sterilization failures.  

 

2. Sterilization Efficacy and Microbial Kill Rate  

Microbial Kill Kinetics  

When dryness is insufficient (e.g., 85%), the increased moisture content in steam lowers the actual temperature inside the sterilizer below the set value (e.g., 121°C may drop to 115°C). This requires extending sterilization time to 1.5 times the original plan to achieve the same kill rate (meeting the Fvalue standard).  

 

Spore Inactivation Challenge  

Dry steam at 121°C can inactivate Geobacillus stearothermophilus spores (D-value = 1.5-2.0 minutes) within 15 minutes, whereas wet steam requires over 25 minutes to achieve the same effect.  

 

3. Operational Risks and Equipment Maintenance  

Condensation Issues  

Wet steam tends to form condensation on sterilized items, leading to wet packs (moisture content >6%), which increases the risk of recontamination and necessitates re-sterilization. Condensation buildup can also corrode sterilizer walls and piping, shortening equipment lifespan.  

 

Energy Consumption  

The low thermal efficiency of wet steam increases energy consumption by 20%-30% for the same sterilization outcome.  

 

4. Measures to Optimize Steam Dryness  

Equipment Improvements  

- Install steam separators to remove entrained liquid water, ensuring dryness 97%.  

- Use steam quality monitors for real-time dryness detection to maintain process stability.  

 

Operational Protocols  

- Preheat pipelines thoroughly before sterilization and drain condensation to prevent wet steam from entering the chamber.  

- Regularly maintain boilers and steam generators to prevent scaling, which can increase steam moisture.


Your cookie settings.

Personalised experiences at full control.

This website uses cookies and similar technologies (“cookies”). Subject to your consent, will use analytical cookies to track which content interests you, and marketing cookies to display interest-based advertising. We use third-party providers for these measures, who may also use the data for their own purposes. 

You give your consent by clicking "Accept all" or by applying your individual settings. Your data may then also be processed in third countries outside the EU, such as the US, which do not have a corresponding level of data protection and where, in particular, access by local authorities may not be effectively prevented. You can revoke your consent with immediate effect at any time. If you click on "Reject all", only strictly necessary cookies will be used.